How to Calibrate a Ribbon Sensor on a Badgy200
If your Badgy200 card printer is not properly recognizing your ribbon, either by not recognizing the type of ribbon or acknowledging that a ribbon has been replaced, you will need to calibrate the ribbon sensors on your Badgy200 printer.
Please note, a YMCKO ribbon must be installed to run the calibration.
Calibrate Ribbon Sensor
- Open the Badgy print Center. Double Click on the Badgy200.
- Click on the maintenance tab, then Printer commands prompt.
- Type in the command Sa;c to calibrate the color sensor. Type in the command Sa;r to calibrate the ribbon rotation sensor. Commands must be entered exact.
If you continue to experience problems, please contact AlphaCard technical support.